MNMR Center

Spectrometer Information

Use the tabs below to view specifications of each instrument

  • 9001
  • 8501
  • 7002
  • 7001
  • 6002
  • 6001
  • 700SS
  • 600SS

Bruker 900 MHz Spectrometer

Description:

The Bruker 9001 has two probes of 1H sensitivity 11,000:1. It is used to collect data for NMR study of proteins, nucleic acids, and small molecules for high sensitivity.

 

CryoProbe 5mm TCI probe of Z-Gradient:

1H S/N of sample 2mM Sucrose - 1,650:1
1H S/N of sample 0.1% EB - 11,000:1
13C S/N of sample ASTM - 2,100:1
VT - 0°C to 80°C

 

Room temp. 5mm TXI probe of Z-Gradient:

1H S/N of sample 2mM Sucrose - 570:1
1H S/N of sample 0.1% EB - 3,000:1
13C S/N of sample ASTM - 380:1
VT - -150°C to 150°C

9001 spectrometer

Bruker 850 MHz Spectrometer

Description:

The Bruker 8501 has two probes of 1H sensitivity 10,000:1and is used to collect data for NMR study of proteins, nucleic acids, and small molecules for high sensitivity.

 

CryoProbe 5mm TCI probe of Z-Gradient:

1H S/N of sample 2mM Sucrose - 1,600:1
1H S/N of sample 0.1% EB - 10,000:1
13C S/N of sample ASTM - 1,800:1
VT - 0°C to 80°C

 

Room temp. 5mm TXI probe of Z-Gradient:

1H S/N of sample Sucrose - 570:1
1H S/N of sample 0.1% EB - 2,500:1
VT - -150°C to 150°C

8501 spectrometer

Bruker 700 MHz Spectrometer

Description:

The Bruker 7002 is one of two 700 MHz spectrometers in the Center. This spectrometer is used for the study of metabonomics (sample volume as little as 30 micro liters) and tissue samples.

 

CryoProbe 1.7mm TCI probe of Z-Gradient:

1H S/N of sample 2mM Sucrose - 380:1
1H S/N of sample 0.1% EB- 1,600:1
13C S/N of sample ASTM - 100:1
VT - 0°C to 80°C

 

4mm HRMAS probe of Z-Gradient:

MAS speed limit - 15 kHz; VT - -80°C to 80°C

 
7002 spectrometer

Bruker 700 MHz Spectrometer

Description:

The Bruker 7001 has three probes of 1H sensitivity up to 7,800:1, used to collect data for NMR study of proteins, nucleic acids, and small molecules for high sensitivity.

 

CryoProbe 5mm TCI probe of Z-Gradient:

1H S/N of sample 2mM Sucrose - 1,130:1
1H S/N of sample 0.1% EB - 7,800:1
13C S/N of sample ASTM - 1,400:1
VT - 0°C to 80°C

 

CryoProbe 5mm TXI probe of Z-Gradient:

1H S/N of sample 2mM Sucrose - 800:1
1H S/N of sample 0.1% EB - 5,300:1
13C S/N of sample ASTM - 380:1
VT - 0°C to 80°C

 

Room temp. 5mm TXI probe of Z-Gradient:

1H S/N of sample 0.1% EB - 1,700:1
VT - -150°C to 150°C

 

7001 spectrometer

Agilent/Varian 600 MHz Spectrometer

Description:

The Agilent/Varian 6002 spectrometer has two probes. This is best suited to collect data for study of metabolomics and small molecules.

 

Room temp. 5mm HCN probe of Z-Gradient:

1H S/N of sample 0.1% EB - 1,200:1
VT - -80°C to 120°C

 

Room temp. 5mm BB probe of Z-Gradient:

1H S/N of sample 0.1% EB - 480:1
13C S/N- 330:1
15N S/N- 35:1
13P S/N- 180:1
19F S/N- 480:1
VT - -80°C to 120°C

6002 spectrometer

Agilent/Varian 600 MHz Spectrometer

Description:

The Agilent/Varian 6001 has two probes of 1H sensitivity 6,500:1, used for NMR study of proteins, nucleic acids, and small molecules for high sensitivity.

 

Cold HCN probe of Z-Gradient (13C enhanced):

1H S/N of sample 2mM Sucrose - 650:1
1H S/N of sample 0.1% EB - 6,500:1
13C S/N of sample ASTM - 630:1
VT - 0°C to 80°C

 

Room temp. 5mm HCN probe of Z-Gradient:

1H S/N of sample 0.1% EB - 1,200:1
VT - -80°C to 120°C

 

6001 spectrometer

Agilent/Varian 700 MHz Spectrometer

Description:

The Agilent/Varian 700SS has six probes to collect data for the study of solid state samples.

 

3.2mm Varian HXY BioMAS: 1H/31P-13C/13C-15N

 

4.0mm Varian HXY T3 MAS: 1H/31P-13C/13C-15N

 

Varian HX BioStatic rectangular coil: 1H/31P-15N for glass plates (300uL)

 

Varian HXY BiCell round coil: 1H/31P-13C/13C-15N for bicelles (150uL)

 

MagLab (Tallahasse,FL) HN rectangular coil: 1H/15N for glass plate samples (300uL)

 

MagLab HX round coil: 1H/15N or 1H/13C for bicelles (150uL)

700SS spectrometer

Agilent/Varian 600 MHz Spectrometer


Description:

The Agilent/Varian 600SS spectrometer has two probes to collect data for solid state studies.

 

3.2mm Varian HXY BioMas:

1H/31P-13C/13C-15N

 

Varian HX BioStatic rectangular coil:

1H/31P-15N for glass plates (300uL)

600SS spectrometer

Go to Top