Instrumentation

There are 8 spectrometers within the Minnesota NMR Center, available for use by both internal and external researchers. If you prefer, one of our specialized managers can operate the spectrometer for you, at a rate of $100/hour. More information on services available here. Training is also available. Visit our Training page for more details.

Spectrometer Rates, effective September 1, 2017

SpectrometerAcademic Rate/HourCommercial Rate/Hour
900 MHz (9001)$12.00$250.00
850 MHz (8501)$11.00$200.00
700 MHz (7002)$8.00$100.00
700 MHz (7001)$7.50$100.00
700 MHz Solid-State (700SS)$8.50$50.00
600 MHz (6002)$8.00$50.00
600 MHz (6001)$8.00$50.00
600 MHz Solid State (600SS)$7.50$30.00

Bruker 900 MHz (9001)

Bruker 900 MHz (9001)

9001.jpg

The Bruker 900-MHz Avance III (9001), equipped with a 5-mm triple resonance cryoprobe, is used for NMR study of proteins, nucleic acids, and small molecules requiring high sensitivity.

CryoProbe 5-mm TCI (1H, 13C, 15N, 2H) w/ Z-gradient

1H S:N (0.1% ethylbenzene) - 11,000:1
1H S:N w/ H2O suppression (2mM sucrose) - 1,650:1
13C S:N (ASTM) - 2,100:1
VT - 0°C to 80°C

Room temp. 5-mm TXI probe w/ Z-gradient

1H S:N (0.1% ethylbenzene) - 3,000:1
1H S:N w/ H2O suppression (2mM sucrose) - 570:1
13C S:N (ASTM) - 380:1
VT - -150°C to 150°C

Bruker 850 MHz (8501)

Bruker 850 MHz (8501)

8501.jpeg

The Bruker 850-MHz Avance III (8501), equipped with a 5-mm triple resonance cryoprobe, is used for NMR study of proteins, nucleic acids, and small molecules requiring high sensitivity.

CryoProbe 5-mm TCI (1H, 13C, 15N, 2H) w/ Z-gradient

1H S:N (0.1% ethylbenzene) - 10,000:1
1H S:N w/ H2O suppression (2mM sucrose) - 1,600:1
13C S:N (ASTM) - 1,800:1
VT - 0°C to 80°C

Room temp. 5-mm TXI probe w/ Z-gradient

1H S:N (0.1% ethylbenzene) - 2,500:1
1H S:N w/ H2O suppression (2mM sucrose) - 570:1
VT - -150°C to 150°C

Bruker 700 MHz (7002)

Bruker 700 MHz (7002)

7002.jpg

The Bruker 700-MHz Avance III (7002) can be used for solution or solid-state NMR.  It is equipped with a cooled SampleJet sample changer (480 samples) for maximum stability in high-throughput studies and a 1.7-mm triple resonance cryoprobe (30 microliter sample volume), in addition to a variety of solid-state probes.

CryoProbe 1.7-mm TCI (1H, 13C, 15N, 2H) w/ Z-gradient

1H S:N (0.1% ethylbenzene) - 1,600:1
1H S:N w/ H2O suppression (2mM sucrose) - 380:1
13C S:N (ASTM) - 100:1
VT - 0°C to 80°C

4-mm HRMAS probe w/ Z-gradient

MAS speed limit - 15 kHz
VT - -80°C to 80°C

3.2mm ssMAS probe

MAS speed limit - 24 kHz
VT - -80ºC to 80°C

Bruker 700 MHz (7001)

Bruker 700 MHz (7001)

7001.jpg

The Bruker 700-MHz Avance (7001) is equipped with a 5-mm triple resonance cryoprobe and a cooled SampleJet sample changer (480 samples) for maximum stability in high-throughput applications.  It is used for NMR study of proteins, nucleic acids, and small molecules requiring high sensitivity.

CryoProbe 5-mm TXI (1H, 13C, 15N, 2H) w/ Z-gradient

1H S:N (0.1% ethylbenzene) - 5,000:1
1H S:N w/ H2O suppression (2mM sucrose) - 900:1
13C S:N (ASTM) - 380:1
VT - 0°C to 80°C

Room temp. 5-mm TXI probe w/ Z-gradient

1H S:N (0.1% ethylbenzene) - 1,700:1
VT - -150°C to 150°C

Agilent/Varian 700 MHz (700SS, Solid-State)

Agilent/Varian 700 MHz (700SS, Solid-State)

700ss.jpg

The Agilent/Varian 700SS has six probes to collect data for the study of solid state samples.

  • 3.2mm Varian HXY BioMAS: 1H/31P-13C/13C-15N
  • 4.0mm Varian HXY T3 MAS: 1H/31P-13C/13C-15N
  • Varian HX BioStatic rectangular coil: 1H/31P-15N for glass plates (300uL)
  • Varian HXY BiCell round coil: 1H/31P-13C/13C-15N for bicelles (150uL)
  • MagLab (Tallahasse,FL) HN rectangular coil: 1H/15N for glass plate samples (300uL)
  • MagLab HX round coil: 1H/15N or 1H/13C for bicelles (150uL)

Bruker 600 MHz (6002)

Bruker 600 MHz (6002)

6002.jpegThe Bruker 600-MHz Avance NEO (6002), equipped with a 5-mm triple resonance cryoprobe, is used for NMR study of proteins, nucleic acids, polymers, and small molecules.

CryoProbe 5-mm TCI (1H/19F, 13C, 15N, 2H) w/ Z-gradient

1H S:N (0.1% ethylbenzene) - 8,300:1
1H S:N w/ H2O suppression (2mM sucrose) - 1,200:1
13C S:N (ASTM) - 1,400:1
19F S:N (TFT) - 7,400:1
VT - 0°C to 80°C

Bruker 600 MHz (6001)

Bruker 600 MHz (6001)

6001

The Bruker 600-MHz Avance NEO (6001), equipped with a 5-mm triple resonance cryoprobe, is used for NMR study of proteins, nucleic acids, polymers, and small molecules.

CryoProbe 5-mm TCI (1H, 13C, 15N, 2H) w/ Z-gradient

1H S:N (0.1% ethylbenzene) - 8,300:1
1H S:N w/ H2O suppression (2mM sucrose) - 1,200:1
13C S:N (ASTM) - 1,500:1
VT - 0°C to 80°C

Agilent/Varian 600 MHz (600SS, Solid-State)

Agilent/Varian 600 MHz (600SS, Solid-State)

600ss.jpeg

The Agilent/Varian 600SS spectrometer has two probes to collect data for solid state studies.

  • 3.2mm Varian HXY BioMas: 1H/31P-13C/13C-15N
  • Varian HX BioStatic rectangular coil: 1H/31P-15N for glass plates (300uL)

Publication Acknowledgement